      Why when I apply an orientation correction, the graph suddenly shows a displacement?

      There is a bug in the correction routine. The bug bites when the 0 and 180 values for a particular depth are exactly the same. For example, the bug will bite when the A0 value and the A180 value at 9.5 meters are both 14. The bug doesn’t bit often because usually the numbers have a different sign and a different value.

      The workaround is simple. Use DMM to edit the data file. Change one of the values by 1 unit. For example, change 14 to 15. That makes the problem disappear and does not affect the accuracy of the data.

      What’s the difference between English and metric templates?

      These templates control the way DigiPro1 processes your inclinometer readings. Use English templates if you have an English-unit inclinometer probe. Use metric templates if you have a metric unit probe. If you need “metric” output from your English-unit inclinometer system, make the conversion in the “Data Units” tab.

      How do I set paper size to A4?

      Click on “file,” then “options and defaults,” then “page setup.” Choose A4 from the drop-down list. Then click OK. Now, just to ensure that the new default isn’t lost, click on “file” and choose “exit.” Click on “yes” to the “save DigiPro1 defaults” question. Now restart DigiPro1. All new reports will be started with A4 size.

      Why doesn’t DigiPro1 doesn’t show all of my depths?

      Short Answer: You’ve probably used an English template to create a graph from metric data. Try using a metric template and see if the results are are better. If this doesn’t help, see the long answer below.

      Long Answer: A “flag” is stored with each survey. The flag tells DigiPro1 that the survey is either English or Metric. Sometimes, especially with manually entered data, the English/metric information is missing. When this happens, DigiPro1 defaults to English units. To set the flag to metric, close the graph. The “installations and reports” dialog reappears. Right-click on the installation and choose “properties” from the pop-up menu. Change the radio button to “metric” and click OK.

      How do you make time-displacement graphs?

      Select the installation, then click “New” and choose “Sample Time – Displacement Graph.” DigiPro1 then plots a graph for the A axis using a single zone. To specify the zones you want, click on the graph to call up the report properties dialog. Then click on the new “Zones” tab. You can specify up to five zones, choosing a start depth (shallower) and a stop depth (deeper) for each zone. You will probably want to click on the “Datasets” tab to select more datasets.

      I can’t see the yellow lines when I print. Is there a way to change colors?

      Click on “File” – “Options and Defaults” – “Preferences.” You’ll see a band of 8 colors. Click on the color you want to change and choose a different color from the pop-up color dialog. The change takes effect when you close your report and open it up again. Later, when you quit DigiPro1, be sure to say “yes” to the prompt “Save DigiPro1 Defaults.”

      How do you specify the initial dataset?

      Just right-click on the dataset that you want to mark as initial, choose “Mark as initial” from the pop-up menu, and click apply. All datasets earlier than the initial are then ignored.

      If we add casing to the top of our inclinometer installation, can DigiPro1 handle this?

      Yes, but you’ll have to do some minor editing in DMM. 

      Why does Error 3051 appear when I try to open a database?

      Check to see if the file is marked “Read-Only.” Using Windows Explorer, select the database, then click the right mouse button and choose Properties. You can find file attributes at the bottom of the properties dialog. Remove the checkmark from Read Only. You may have to click the box more than once. Now DMM and DigiPro1 can open the database.

      Do you have any instructions for installing horizontal casing?

      The manual for the horizontal probe provides some instructions for installing horizontal casing in soil. Most of these instructions will apply to your application as well. For example,

      • The traversing probe requires 70 or 85 mm (2.75 or 3.34″) casing.
      • One set of grooves must be vertical. This is essential. Check for this during installation.
      • Telescoping sections are not recommended or needed.
      • Casing should be flat and straight. It’s ok to build in an inclination for drainage.
      • A pull cable is required to pull the probe to the far end. Install the cable as you assemble the casing.

      For your application, also consider this:

      • Support the casing well, at least every 5 feet and under couplings. Tie the casing to the supports so that it does not float upwards when the concrete is poured. Wrap the joints with tape to make sure grout does not enter.
      • Remember that the casing is plastic and weaker than steel. Avoid pouring concrete directly onto the casing.
      • The casing is plastic and will start deforming if temperatures exceed 82° C ( 190° F). If the ambient temperature is above 26° C (80° F), you may need to fill the casing with cool water at the same time as you are pouring concrete. Make sure that the casing is supported by the concrete before filling it. There are other ways to avoid high temperatures as well, such as icing the concrete.

      Why does an error 11 appears when I try to create a graph using DigiPro1?

      Error 11 is caused by a problem in your data file, usually a missing instrument constant. Open your database with DMM. Click on an installation, then click the survey list tab. Check every survey has an instrument constant. The survey that is missing the constant is causing the problem. Use Edit/Add to enter the constant.

      Instalación general

      DigiPro2 es distribuido como una versión de prueba. La versión de prueba en sus primeros 45 días, después de la instalación, trabaja con todas sus características y ventajas activadas. Después de este periodo, la versión avanzada se transforma a la versión básica.

      La versión básica de DigiPro2 es gratis y permite el manejo de bases de datos y crear gráficos primarios. DigiPro2 puede ser convertido en cualquier momento a la versión avanzada con solo ingresar una clave valida.

      Claves para activar DigiPro2 avanzado deben ser compradas. Para adquirir una licencia, contacte a DGSI o cualquiera de sus distribuidores. Cuando se compra una licencia nosotros ingresamos información de su compañía en el ordenador de licencias de DGSI y este emite un código que genera la clave. Después de esto le enviaremos el código de la clave por correo electrónico. Si usted compra una licencia de un distribuidor, la clave será enviad al distribuidor y ellos se la enviaran a usuario final.

      Los paso que siguen, explican como ingresa y manejar la clave.

      Como Instalar la Clave de la LicenciaDigiPro2 Clave de la LicenciaPaso 1: Comience DigiPro2 y haga clic en “License”. Después, DigiPro2 mostrará el cuadro de diálogo
      Paso 2: Copie la clave del correo electrónico que recibió y péquelo en el cuadro “Licence Key”. De aquí haga clic en “Install”
      Paso 3: DigiPro2 valida el código de la clave a través del Internet. Cuando la validación se completa usted vera el mensaje siguiente ” License Installed” en el área llamada “License Status.”Hacer clic en “Close” para completar el proceso.

      Como Trasladar la Licencia

      Si es necesario trasladar la licencia del ordenador original a uno nuevo hay que seguir los siguientes pasos:

      1. Estar seguro que usted tiene la clave que se le envió originalmente mediante correo electrónico.
      2. Comience DigiPro2 en el ordenador antiguo.
      3. Hacer Clic en “License”; en “Unistall License” escoja “Yes”. Al hacer esto, DigiPro2 se convierte a la versión básica en este ordenador.
      4. Instale DigiPro2 en el ordenador nuevo. Pegue la clave en “License Key” de la misma forma que se explica arriba.