      My surveys are incorrectly marked “English” but I have a metric system (or my surveys are incorrectly marked “metric” but I have an English system). What’s going on?

      There are two ways this can happen. (1) You are accidentally telling the DataMate that you have a metric (or English) system, or (2) the installation information in the DataMate is incorrect.

      Case 1: When you start a survey using the DataMate, you choose an installation and then you must step through the installation parameters by pressing the Enter key. If you press the down arrow by mistake, you’ll change the value of that parameter. For example, if you press the down arrow at the Units prompt, you’ll change the value from English to metric or vice versa. It seems logical that you could press the down arrow to scroll through the readings, but you can’t. Always press the Enter key to step through the parameters.

      Case 2: You should check that installation information in the DataMate is correct. Switch on the DataMate. Choose Read to display the Read menu. Then choose Installations. Scroll through the list of installations and choose the one that is causing problems. Then step through the installation parameters until you see Units. Set this to English or metric, depending on the type of probe that you have. Press Enter to see the next parameter, Ins Constant. Set this to 20000 for English probes or 25000 to metric probes. Also, check your DMM database to see that probe type and instrument constant are set properly for that installation.

      Fixing Incorrectly Recorded Data: If you recorded an English probe with a metric setting in the DataMate, or you recorded a metric probe with an English setting in the DataMate, your data values are not correct. There are three solutions.

      1. In DMM, edit each value in the affected survey.

      2. In DMM, edit the

      Apply a sensitivity correction in DigiPro.

      3. Change the Probe Constant for that You can edit each value in the survey and keep , you can or you can change the instrument constant for each affected survey so that computed deviations and displacements will be correct. Here are instructions:

      1. Open the project database, choose the appropriate installation.
      2. Choose the appropriate survey and click the Edit Add button.
      3. Change the value in the “probe constant” field as follows:
      PROBE UNITSACCIDENTAL SETTING FOR THIS SURVEYUSE THIS PROBE CONSTANT IN DMM
      EnglishMetric35000 (If your DMM does not accept this value, use the sensitivity correction in DigiPro)
      MetricEnglish14286
      PROBE UNITSACCIDENTAL SETTING FOR THIS SURVEYUSE THIS SENSITIVITY CORRECTION IN DIGIPRO
      EnglishMetric1.75
      MetricEnglish0.5714

      Why if most of my surveys are marked Full Set = “T.” do I sometimes see a survey marked “Full Set = F.”?

      The full set flag tells DigiPro how to process the data. A T value indicates a normal two-pass survey, in which each depth has a 0 reading and a 180 reading). The software then combines the two readings and divides by 2. An F value tells the software that there is only a 0 reading and therefore no combining or dividing takes place.

      If you end your normal two-pass survey by choosing “Done,” the flag is set to T. However, if you end your survey by pressing Esc, the flag is set to False, even though all the data are present.

      I’m trying to import GTilt files, but DMM shows an error message – something about a field cannot have a zero value.

      Your G-Tilt file may be missing a sensor (probe) serial number.

      G-Tilt marks the start of every survey with a *. Count six lines down from the *. You should see a serial number or at least some sort of text. Go through each of your G-Tilt files. Look for the * and count six lines down. Enter the sensor serial number in that 6th line (or just type 12345). Look for additional *s in the file and check that the sixth line after each one has a serial number. You can put any text tha fIf you have multiple surveys in the file, you’ll have to make multiple entries.

      *                         —- marks the start of a survey
      070685             —- survey date
      1327                 —- time hhmm
      1400                 —- not used by DMM
      Top of clamp   —- not used by DMM
      31.2                  —- not used by DMM
      12345               —- sensor serial number. line must not be empty

      Digipro Update Information

      Digipro for Windows has been replaced by DigiPro2.

      DigiPro2 replaces both DigiPro1 and DMM.

      Why did you replace the original DigiPro for Windows?

      DigiPro and DMM were written in Microsoft VB6. Microsoft discontinued support for VB6 development tools some time ago, making bug fixes and further development nearly impossible. In addition, Microsoft warned that Windows 7 would not support for the VB6 runtime. In fact, they relented and included some minimal VB6 runtime support in Win7, but it is certain that DigiPro and DMM will stop running in some future update of Windows.

      For that reason, we developed DigiPro2 to replace both DMM and DigiPro. Development took over a year, but it is written in C# and is guaranteed to run on future updates of Windows. We encourage everyone to migrate to DigiPro2 sooner rather than later.

      You can download a trial version of DigiPro2. It runs in advanced mode for 45 days, then it reverts to a basic version that you can continue to use.

      I want to include DigiPro graphs in my report. How can I export the graphs?

      There is no direct way to export graphs. (Note: this is a built -in feature of DigiPro2).

      With DigiPro1, you can print graphs to a software “image” printer. The image printer creates a graphic file, which can be emailed, placed in Word, etc.

      We have good results with “Zan Image Printer.” It is inexpensive, very convenient to use, and widely available on the internet. We are using Zan to print the DigiPro graphs that we upload to Atlas, our web-based monitoring system.

      You can also use a screen capture program, such as “HyperSnap” or “Snag It”. However, screen captures are low resolution, good enough for the computer monitor, but not good enough for printing.

      Why does DigiPro1 consistently plots graphs with one extra depth?

      Think of a measuring tape. The tape starts at zero. You don’t record the zero, but it is the origin of your length measurements. Note that the plotted value for the extra depth is always zero.

      If your inclinometer installation is 12m deep, and you take readings at half-meter intervals, you will take 23 readings. Your bottom reading is taken at 11.5 meters, but the origin of that interval is at 12 meters. When DigiPro draws the plot, it establishes a zero at 12 meters.

      Is sand backfill as good as grout backfill?

      No. Sand backfill has two negatives: it settles, and it washes away.

      One west coast department of transportation has studied this. According to them, sand backfill settles for about a month after installation, unless it is well tamped. Unfortunately, it is very difficult to tamp effectively in a small diameter borehole. This settlement means that a good initial reading cannot be obtained for a month after installation. If you obtain the reading earlier, you are likely to detect only movement of the casing within the borehole, rather than movement of the ground.

      In ground that has faults sand is sometimes used because grout leaks out the fault. Over time, however, the sand will wash away. Then the casing will move within the borehole and all readings are suspect.

      Finally, if there are gaps in backfill (and this goes for grout as well as sand), you’ll find that it takes longer for the inclinometer probe to stabilize. Evidently, movement of the probe in the casing can set up some vibrations.

      Why doesn’t DigiPro1 plot displacements at the same depths that I recorded them?

      This is the result of “auto depth-adjustment. With auto-depth adjustment turned on, DigiPro1 correctly plot data points at the top (or bottom) of the measurement interval. Auto-depth is turned on by default.

      Why is an adjustment provided? Depth marks on Digitilt control cable are measured from the middle of the inclinometer probe, but deviations and displacements are calculated for the top (or bottom) of an interval.

      Metric example: The depth stored with the inclinometer reading is the cable depth of 20 meters, but the top of the interval is actually at 19.75 meters. With auto-depth adjust turned on, the plotted point will be placed correctly on the graph at 19.75 meters, rather than at the cable depth of 20 meters.

      English example: The depth stored with the inclinometer reading is the cable depth of 60 feet, but the top of the interval is actually at 59 feet. With auto-depth adjust turned on, the plotted point will be placed on the graph at 59 feet, rather than at the cable depth of 60 feet.

      What is Depth Offset / Elevation Offset and how should I use it?

      During a survey, the depth of the probe is controlled by aligning depth marks on the control cable to a reference, such as the top of the casing or the top of a pulley assembly. If you want depths on the plot to represent depths below ground surface, then you should enter an offset to adjust the reference to ground surface. That is the purpose of the offset field.

      For Depth Offset, enter height of casing above ground level + height of pulley (1 ft or 0.3m)

      Metric Example: Suppose the top of the casing is 0.5 meters above ground level. You use a pulley and index the cable to the top of the pulley (add 0.3m). Enter 0.8 m as the depth offset. Labels will show depth below ground surface. .

      English Example: The top of the casing is 14 inches (1.17 ft.) above ground level. The pulley assembly adds 1 foot. Enter 2.17 feet for the depth offset. Labels will show depth below ground surface. .

      For Elevation Offset, enter ground elevation + casing height + pulley height

      Metric Example: Ground elevation is 200 meters above sea level. The top of the casing is 0.4 meters above ground level. The pulley assembly adds 0.3 meters. Enter 200.7 meters for the elevation offset. Labels will be referenced to ground elevation.

      English Example: Ground elevation is 1200 feet above sea level.
      The top of the casing is 1.5 feet above ground level. The pulley assembly adds 1 foot. Enter 1202.5 feet for the elevation offset. Labels will be referenced to ground elevation.

      How are data points calculated for the time displacement graph? 

      The time displacement graph requires a single value for each zone. To do this, we first calculate cumulative displacement values for every depth. Then, for each zone, we subtract the value at the bottom depth of the zone from the value at the top depth of the zone. The sign is changed to accommodate top or bottom reference. So the zero point for the plotted value is the value at the bottom of the zone.

      How can I print the values used in a graph?

      In DigiPro1, first plot your data. click on one of the plots to display the properties. Right click in the ‘Data Sets’ window, and select ‘Print plotted data’ from the popup menu. This will print the calculated data used to make the plot.