What effect does cable length have on the thermistor reading?
The AWG 22 copper wire in signal cable for piezometers has a nominal resistance of 16 ohms per 1000 feet. The round trip of the wire-pair to and from the thermistor effectively doubles this value to 32 ohms.
The thermistor used in our sensors, as well as in those of most other instrument manufacturers, provides a resistance of 3,000 ohms at 25 degrees C. The thermistor is non-linear, however.
At 25 degrees C, 32 ohms represents about 0.24 degree C, and at 10 degrees C, 32 ohms represents about 0.11 degree C. In other words, the effect of cable length has less effect as the temperature decreases. In either case, though, correction for cable length is probably not needed.
If we cut off cable, do we need to adjust the RTD offset?
After installing some VW instruments, we had to cut the cables to connect to the data logger. Do we need to adjust the RTD offset, and if so, how can we do this in the field with all the instruments installed?
For many, this question is not relevant. Our VW piezometers are almost always shipped with thermistors rather than RTDs.
For those with older VW piezometers: T he VW sensor calibration record lists an RTD offset based on the original length of the cable that was attached to your sensor. If you change the length of the cable, you must adjust the offset. Use this table to find the change in offset for a given change in cable length. If you have shortened the cable, subtract the change in offset from the original offset. If you have lengthened the cable, add the change in offset to the original offset. Note that the vibrating wire part of the signal (the pressure measurement) is not affected by cable length.
Also, take a look at the most recent version of the VW piezometer manual for a discussion about whether you really need to go through these steps.
Why are the temperature readings from my piezometer between 120 and 125°C?
You’re probably using the VW Data Recorder and reading a thermistor as an RTD. At the “type” prompt, choose Hz + Thermistor. Then your temperature readings will look normal.
Should I set up the GTecLink nodes or Gateway first?
If you are using a Gateway, you should set up the Gateway first. Once the Gateway is operational, each node should be commissioned near to its final installation location. This will allow you to test the radio signal strength between the node and the Gateway to verify you will have successful communication.
What RTD do you use and what are its characteristics?
Unless you ask for an RTD, we now supply a thermistor.
In previous years, we used the Honeywell TD5A sensor. Here is an Acrobat datasheet listing its calibration characteristics. More general information appears below:
- Sensor Type: 2000 Ohm Silicon RTD
- Temperature Sensing Range: -40 °C to 150 °C (-40 °F to 302 °F)
- Linearity % of Full Scale: ± 0.2%, Typ. -40 °C to +150 °C [-40 °F to 302 °F]
- Supply Voltage: 10 Vdc
- Supply Current (Recommended): 1 mA typ., 2 mA max
If I am not using a DGSI readout, how can I obtain a reading in degrees C from the RTD in your sensors?
Our published instructions assume that users are reading these 2K ohm RTDs using DGSI readouts or the Campbell Scientific CR10X with AVW1 or AVW100 vibrating wire interface.
If you are using another device to read these sensors, obtain a reading in K ohms. Then apply either of the following polynomial coefficients to obtain degrees C:
Solution A: For a range of -10 to +30 C.
Ax2 + Bx + C, where x is the reading in K ohms, and:
A = -23.508334394
B = 227.625006633
C = -341.217356436
Solution B: For a range of -50 to 120 C.
Ax5 + Bx4 + Cx3 + Dx2 + Ex + F, where x is the reading in K ohms, and:
A = 0.95659
B = -12.65962
C = 70.30009
D = -216.32308
E = 483.85218
F = -472.87637
What’s the maximum cable length I can have when using a datalogger to read 4-20mA titanium pressure transducers?
Approximately 240 meters or 800 feet. The datalogger supplies 12 volts, and the transducer requires at least 8 volts. The standard cable used with the 4-20mA transducer has 22 gauge wire so voltage drops about 0.6 volts per 1000 feet.
How do I calculate linear factors (mx + b) using the data on my calibration record?
Please note that we no longer calibrate VW piezometers with this method, however the information is presented here for older piezometers that were calibrated utilizing the old method.
We’ve made an Excel worksheet called: vw-linear-factors-xls. Your calibration record gives frequency readings for 12 pressures. Enter these Hz readings in the frequency column, overwriting existing data. The spreadsheet converts calculates frequency squared readings and then applies Excel slope and offset functions to calculate an m factor and a b factor. These are labelled slope (m) and offset (b). You’ll need WinZip or some other archiving program to unzip the file.
Can you tell me more about piezometers in fully grouted boreholes?
The grout-in method is a faster and easier way to install piezometers. It eliminates the conventional sand filter and the bentonite seal. Instead, the entire borehole is filled with a non-shinking, low permeability grout.
Visit the Theories and Setups page to learn about the grout-in method.
Download the VW Piezometer and Multi-Level VW Piezometer manuals for instructions.
Do I need to have the Gateway connected to the internet when I set it up?
It is not required, but it is helpful. When the Gateway is first powered up, it first checks for a hardwired LAN connection and if it doesn’t find one, then looks for a SIM card. If it does not find either, it goes into standalone mode. This setting can be changed by hard-wiring your PC into the Gateway and changing it or by powering the Gateway off and back on, but it is simpler if it is connected to the network during its initial commissioning.
You mentioned connecting to the Gateway with a PC. Can I use the GTecLink app to commission the Gateway?
No. The GTecLink app is used only to connect with the nodes to commission them or to download data from them. The initial commissioning of the Gateway requires that a PC be connected to the Gateway via the USB port on the Gateway.
You will need a standard CAT 5 or CAT 6 ethernet cable and will connect one end to your PC and the other to the Ethernet to USB adapter that was included in the Gateway package. If your PC does not have an ethernet port, you will need to order an additional Ethernet to USB adapter (DGSI p/n 58820530) and will have an adapter on both ends of the ethernet cable. Once you are connected to the Gateway and it is powered on, you will open an internet browser and type http://169.254.0.1 in the address bar.
The user name will be admin and the password will be VMjG6z. This address, user name and password will be the same for any wired connection to any Gateway. Please note that this is NOT the address, user name or password to access the Gateway remotely.
That information is provided on the Gateway Information sheet that was included with your Gateway and is unique to each Gateway.
What kind of battery does the V-Logger use, and how do I replace it?
The standard battery is a size D lithium battery, and should be replaced using the information detailed here. It is a 3.6v lithium battery. The preferred manufacturer is Xeno Energy. They sell the battery under part number #XL-205F (a secondary manufacturer also makes them: Tadiran, under part number #TL-5930 (TL-5930/S)). They can be purchased from several online retailers, including Amazon and Sears, as well as at battery specialist stores.