Moisture in the connectors is the most common cause. There are three connectors that could be affected: the connectors on the control cable and the connector on the probe. Use an ohmmeter set to the 2 Megaohm range to check. Check cable connectors when the cable is not connected to the probe or the DataMate. Check the resistance from pin to pin and pin to connector body. In each case, the ohmmeter should read overrange (infinite resistance). If you see another reading, there is a problem. A reading cause by moisture usually drifts (does not remain steady). Continuity checks for the conductors should read about 1 ohm per 100 feet of cable. Check the pin to probe body resistance (not pin to pin). The roll pins (pins that go through the body to hold the springs) make a good connection to the body. Again, the reading should be overrange. If you find a problem in the cable connector that plugs into the DataMate, you may be able to disassemble and dry the connector by yourself. See this technote for directions. However, if you find moisture in the cable connector that attaches to the probe or in the connector built into the probe itself, you must send them to the factory for servicing. Send both items. See instructions for returning equipment to Slope Indicator.