July 31, 2024

Choose a formula from the table below:

Vibrating Wire Sensors with AVW200: 
Where Ohms = reading in ohms

Coefficients for Thermistor: A = 1.401e-3     B = 2.377e-4     C = 0.00e0     D = 9.730e-8

Formula for Thermistor: Deg C = 1/(A+B*LN(Ohms) + C*LN(Ohms)^2 + D*LN(Ohms^3)-273.2

Coefficients for RTD: J = -23.50833439     K = 227.625007     L = -341.217356417

Formula for RTD: Deg C = (J*((Ohms/1000)^2)) + (K*(Ohms/1000))+(L)
Vibrating Wire Sensors with AVW1, AVW100, or AVW4: 
Where VT = reading in volts

Coefficients for Thermistor: A=43.089     B=-240.91     C=544.27    D=-611.59    E=378.11     F=-104.78

Formula for Thermistor: Deg C = A*(VT^5) + B*(VT^4) + C*(VT^3) + D*(VT^2) + E*(VT) +

FCoefficients for RTD: A=-435.97     B=1904.4     C=-3442.5     D=2412.6

Formula for RTD: Deg C = A*(VT^3) + B*(VT^2) + C*VT + D
MEMS or EL SC Sensors: 
Where VT = Thermistor reading in volts:

Coefficients: A=9.3219     B=-54.3038     C=131.165     D=-161.2568     E=137.7711     F=-37.7705

Formula: Deg C = A*(VT^5) + B*(VT^4) + C*(VT^3) + D*(VT^2) + E*(VT) + F
4-20mA Sensors: 
Where VT = Thermistor reading in volts:

Coefficients: A=43.089     B=-240.91     C=544.27    D=-611.59     E=378.11     F=-104.78

Formula: Deg C = A*(VT^5) + B*(VT^4) + C*(VT^3) + D*(VT^2) + E*(VT) + F