July 31, 2024

Short Instructions for Reading Twin-Tube Piezometer

  1. Before leaving office, check that the indicator has enough gas.
  2. At the site, turn flow control valve to Vent and zero the pressure gauge.
  3. Turn flow control valve Off.
  4. Wipe off all dirt and moisture from the transducer tubing and connectors.
  5. Connect the quick connect plug to Transducer socket and vent tube to the return flow indicator.
  6. Turn tank control valve On, then turn the regulator knob clockwise until pressure reaches the rating of the gauge. 7. Turn flow control valve On.
  7. Adjust the flow rate knob so that the pressure reading on the main gauge rises at about 1 PSI per second.
  8. Wait until ball in return flow indicator shows a return flow of gas from the transducer, then turn flow control valve Off.
  9. Wait until pressure reading stabilizes, then tap gauge and write down reading.
  10. To verify, briefly turn flow control valve to Vent, then return to On. This changes the pressure reading. Now take a second, verification reading.